Objective:To compare the clinical performance of a real-time PCR assay with the COBAS Amplicor HBV Monitor test for quantitation of HBV DNA in serum samples.Methods:The reference sera of the Chinese National Institute for the Control of Pharmaceutical and Biological Products and the National Center for Clinical Laboratories of China,and 158 clinical serum samples were used in this study.The linearity,accuracy,reproducibility,assay time,and costs of the real-time PCR were evaluated and compared with those of the Cobas Amplicor test.Results:The intra-assay and inter-assay variations of the real-time PCR ranged from 0.3% to 3.8% and 1.4% to 8.1%,respectively.The HBV DNA levels measured by the real-time PCR correlated very well with those obtained with COBAS Amplicor test(r=0.948).The real-time PCR HBV DNA kit was much cheaper and had a wider dynamic range.Conclusions:The real-time PCR assay is an excellenttool for monitoring of HBV DNA levels in patients with chronic hepatitis B.
|